For this work, a peristaltic micropump was fabricated. Actuation of the micropump was accomplished with piezoelectric cantilevers. To date, a minimal number of soft polymer-based micropump designs, have explored the use of piezoelectric materials as actuators. The fluidic channel for the micropump was fabricated using PDMS and soft lithography. A novel and very simple template fabrication process was employed, where the use of a mask and clean room facilities was not required. Replica molding to the template produces both, a channel measuring in height, and a rounded cross-sectional geometry, the latter of which is known to be favorable for complete valve shutoff. Clamps were adhered to the tips of the cantilevers, and used to secure in place aluminum valves. The valves had finely machined tips on one surface. These tips served as contact points for the valve making contact with the PDMS membrane surface, and were used for the purpose of opening and closing the channels. The cantilevers were secured in place with in-house manufactured micropositioners, which were used to position the valves directly over the PDMS channel. The micropump was thoroughly tested where the variables characterized were maximum attainable backpressure, flow rate, valve open/close characteristics, and valve leakage. The effect of the phase difference (, , and ) between the square wave signals delivered to each of the three cantilevers was investigated for flow rate and maximum attainable backpressure. Of the three signal phases, the signal demonstrated the largest flow rate range of 52–575 nL/min (0.1–25 Hz), as well as the highest attainable backpressure value of 36,800 Pa (5.34 psi). The valve shutoff characteristics for this micropump was also examined. Fluorescein was trapped inside the microchannel, where the fluorescent signal was monitored throughout the valves open/close cycle with the aid of an epifluorescent microscope. It was found that the fluorescent signal went to zero with the valve fully closed, supporting the conclusion that the valve completely closes off the channel. Further evidence of this claim was demonstrated by observing the valve leakage characteristics. An electronic pressure sensor was used to collect data for this experiment, where it was found the valve was able to hold off 36,800 Pa (5.34 psi), only loosing 2% of this pressure over 10 minutes. In conclusion, it has been shown this micropump outperforms many existing micropump designs, and is suitable for integration into a variety of both macro, and microdevice platforms. Experiments are currently underway to examine how the flow and valving characteristics change for valves with different tip dimensions. A discussion will also be given for improved fabrication techniques, where injection molding is currently being used as the fabrication method to examine the performance changes associated with different cross-sectional PDMS channel geometries. The end goal for use of this micropump is twofold; 1) integration into a micro-free flow separation device, and 2) integration into a capillary electrophoresis instrument for use in direct-sampling neuroscience experiments.
Skip Nav Destination
Article navigation
Design Of Medical Devices Conference Abstracts
A Soft-Polymer Piezoelectric Bimorph Cantilever-Actuated Peristaltic Micropump
N. J. Graf,
N. J. Graf
University of Minnesota
, Minneapolis, Minnesota, USA
Search for other works by this author on:
M. T. Bower
M. T. Bower
University of Minnesota
, Minneapolis, Minnesota, USA
Search for other works by this author on:
N. J. Graf
University of Minnesota
, Minneapolis, Minnesota, USA
M. T. Bower
University of Minnesota
, Minneapolis, Minnesota, USAJ. Med. Devices. Jun 2009, 3(2): 027536 (1 pages)
Published Online: July 9, 2009
Article history
Published:
July 9, 2009
Citation
Graf, N. J., and Bower, M. T. (July 9, 2009). "A Soft-Polymer Piezoelectric Bimorph Cantilever-Actuated Peristaltic Micropump." ASME. J. Med. Devices. June 2009; 3(2): 027536. https://doi.org/10.1115/1.3147498
Download citation file:
Get Email Alerts
Cited By
Context-Driven Design of a Laparoscopic Instrument Cleaner for Use in Rural Low-Resource Hospitals
J. Med. Devices (March 2025)
Controlled Ice Nucleation With a Sand-PDMS Film Device Enhances Cryopreservation of Mouse Preantral Ovarian Follicles
J. Med. Devices (December 2024)
Review of Blood and Fluid Warming Methods
J. Med. Devices (December 2024)
Related Articles
Development of a Fiberoptic Distributive Shear/Pressure Sensor
J. Med. Devices (June,2009)
Design and Fabrication of a Micropump With Mixing Function for LOC
J. Med. Devices (June,2008)
Effects of Dissimilar Electrode Materials and Electrode Position on DNA Motion During Electrophoresis
J. Nanotechnol. Eng. Med (May,2011)
Electrokinetic Flow Dynamics of Weakly Aggregated λDNA Confined in Nanochannels
J. Fluids Eng (December,2011)
Related Proceedings Papers
Related Chapters
Key Technology of Interactive Music Game Installation based on Arduino
International Conference on Control Engineering and Mechanical Design (CEMD 2017)
Semi-Analytical Model of the Pull-In Behavior of an Electrostatically Actuated Cantilever Microbeam
International Conference on Mechanical and Electrical Technology, 3rd, (ICMET-China 2011), Volumes 1–3
Nanomechanical Cantilever Biosensors: Conceptual Design, Recent Developments, and Practical Implementation
Biomedical Applications of Vibration and Acoustics in Therapy, Bioeffect and Modeling