Medical, therapeutic and technological advancements, including the use of neonatal incubators and the development of neonatal intensive care units (NICUs), have significantly increased the survival of premature and ill infants. However, high levels of noise in the NICU result in numerous adverse health effects, including hearing loss, sleep disturbances and other forms of stress. Even normal levels of ambient noise may be of considerable risk for the most premature infants. It is well documented that the mammalian auditory system is most vulnerable to environmental influences immediately after the time that it first begins to function. In humans, the critical period spans approximately weeks 24–30 of gestation, which corresponds to the age when the most extremely premature infants are now able to survive ex utero. Premature infants are, therefore, at high risk for environmentally-induced hearing loss. Development of techniques that increase the amount of protection against noise-induced hearing loss (NIHL) could significantly improve quality of life, both while neonates are in the NICU, and long term. The long-term goal of our research is to develop a version of an existing active noise cancellation (ANC) system that can be used to reduce sound levels in NICU incubators, in a manner that does not require considerable space. The core component of the ANC system is a carbon nanotube-based transparent actuator, which is controlled by an adaptive controller so that an exact out-of-phase anti-noise can be produced from the actuator (Yu et al, 2005, 2007). The basic principle of the ANC system is to cancel the unwanted primary noise through the introduction of a destructive anti-noise sound. Experimental results showed that a reduction of greater than 15 dB in the primary noise can be achieved by the ANC system (Yu et al. 2007). Ultimately, this transparent actuator could be built into the side of an infant incubator, providing noise protection without adding equipment to the already crowded NICU environment. Before human trials can begin, animal studies must be completed to demonstrate that the ANC system can prevent the hearing loss that results from exposure to incubator noise during the critical period. One complication in animal testing is that individual species respond to different frequency ranges. For example, the human cochlea is most sensitive to sound frequencies between 2 and 5 kHz, while mice respond best between 8 and 16 kHz. It was hypothesized that a frequency translation based on the cochlear frequency/place relationship could be used to convert incubator noise into an appropriate stimulus for testing of the ANC in mice. Neonatal mice were exposed to untranslated incubator noise (IN) or frequency-shifted incubator noise (FSIN) during the critical period, and hearing sensitivity was measured following the noise exposure. IN had no effect on acoustic thresholds, but FSIN caused a moderately severe (60–70 dB) high frequency hearing loss in all mice tested. Based on these data, the FSIN stimulus represents the first accurate model of neonatal noise-induced hearing loss. Future experiments will use this model to test the ability of the ANC system to protect against NIHL.
Skip Nav Destination
Article navigation
Design Of Medical Devices Conference Abstracts
Development of an Animal Model to Test an Active Noise Cancellation System for Infant Incubators
Z. Tridane,
Z. Tridane
University of Minnesota Duluth
, USA
Search for other works by this author on:
I. M. Hayee,
I. M. Hayee
University of Minnesota Duluth
, USA
Search for other works by this author on:
G. Nordehn,
G. Nordehn
University of Minnesota Medical School
, Duluth, USA
Search for other works by this author on:
J. Fitzakerley
J. Fitzakerley
University of Minnesota Medical School
, Duluth, USA
Search for other works by this author on:
Z. Tridane
University of Minnesota Duluth
, USA
X. Yu
University of Minnesota Duluth
, USA
I. M. Hayee
University of Minnesota Duluth
, USA
G. Nordehn
University of Minnesota Medical School
, Duluth, USA
J. Fitzakerley
University of Minnesota Medical School
, Duluth, USAJ. Med. Devices. Jun 2009, 3(2): 027530 (1 pages)
Published Online: July 9, 2009
Article history
Published:
July 9, 2009
Citation
Tridane, Z., Yu, X., Hayee, I. M., Nordehn, G., and Fitzakerley, J. (July 9, 2009). "Development of an Animal Model to Test an Active Noise Cancellation System for Infant Incubators." ASME. J. Med. Devices. June 2009; 3(2): 027530. https://doi.org/10.1115/1.3147384
Download citation file:
698
Views
Get Email Alerts
Cited By
Related Articles
Automatic Vibrotactile Device for Interruption of Apnea in Premature Infants
J. Med. Devices (June,2010)
Mechanical Design for Modularity of Laparoscopic Tools
J. Med. Devices (June,2008)
Development of a Database for Global Health Medical Devices
J. Med. Devices (June,2011)
Novel Design for Jaw-Thrust and Head Immobilization Device and Its Successful Testing Using Human Simulator
J. Med. Devices (June,2010)
Related Proceedings Papers
Related Chapters
Toward More Effective Evaluation and Control of Airport Noise
Community Noise
mDFA Human Empirical Results
Modified Detrended Fluctuation Analysis (mDFA)
A Human Reliability-Centered Approach to the Development of Job Aids for Reviewers of Medical Devices That Use Radiological Byproduct Materials (PSAM-0299)
Proceedings of the Eighth International Conference on Probabilistic Safety Assessment & Management (PSAM)