Objective: To design an external fixator for femoral fracture reduction with improved angulation and adjustability, reduced bulk, and comparative stiffness. Background: The use of external fixator for femoral shaft fixators in children has been a device of choice. The use though satisfactory, the occurrences of wire∕pin tract infections has been widely reported. Moreover, The use of Ilizarov external fixator in adults has been limited due to its bulk. Hence there is a need for improvement on the current external fixators for femoral fractures. Methods: The finite element models of four-ring Ilizarov and hybrid two-ring fixators were developed. These designs were improved upon by incorporating a “modular sliding joint design,” and by reducing the bulk by reducing the number of rings and adding guide plates. Wire length optimization feature was also added in the design. Axial stiffness of the new design was compared with the hybrid design using finite element analysis (FEA). Results and Discussion: The new design has reduced bulk at the proximal region allowing its application in adults. The axial stiffness of the proposed design was found to be comparable with the two-ring hybrid design, as determined from FEA. Sliding joint design allows reduced inventory, quick assembly, and improved angulation over current designs. Wire length optimization may reduce the occurrences of wire tract infections.
Skip Nav Destination
Article navigation
Design Of Medical Devices Conference Abstracts
A New External Fixator Design for Femoral Fracture Reduction
Michael J. Cucinotta,
Michael J. Cucinotta
The College of New Jersey
, Ewing, NJ 08628 USA
Search for other works by this author on:
Courtney E. LeBlon,
Courtney E. LeBlon
The College of New Jersey
, Ewing, NJ 08628 USA
Search for other works by this author on:
Nima Rahimi,
Nima Rahimi
The College of New Jersey
, Ewing, NJ 08628 USA
Search for other works by this author on:
Manish Paliwal
Manish Paliwal
The College of New Jersey
, Ewing, NJ 08628 USA
Search for other works by this author on:
Michael J. Cucinotta
The College of New Jersey
, Ewing, NJ 08628 USA
Courtney E. LeBlon
The College of New Jersey
, Ewing, NJ 08628 USA
Nima Rahimi
The College of New Jersey
, Ewing, NJ 08628 USA
Manish Paliwal
The College of New Jersey
, Ewing, NJ 08628 USAJ. Med. Devices. Jun 2008, 2(2): 027564 (1 pages)
Published Online: July 2, 2008
Article history
Published:
July 2, 2008
Citation
Cucinotta, M. J., LeBlon, C. E., Rahimi, N., and Paliwal, M. (July 2, 2008). "A New External Fixator Design for Femoral Fracture Reduction." ASME. J. Med. Devices. June 2008; 2(2): 027564. https://doi.org/10.1115/1.2932718
Download citation file:
Get Email Alerts
Cited By
Context-Driven Design of a Laparoscopic Instrument Cleaner for Use in Rural Low-Resource Hospitals
J. Med. Devices (March 2025)
Controlled Ice Nucleation With a Sand-PDMS Film Device Enhances Cryopreservation of Mouse Preantral Ovarian Follicles
J. Med. Devices (December 2024)
Review of Blood and Fluid Warming Methods
J. Med. Devices (December 2024)
Related Articles
Design of a Dynamic Stabilization Spine Implant
J. Med. Devices (June,2009)
Synthetic Soft Tissue Characterization of the Mechanical Analogue Lumbar Spine
J. Med. Devices (June,2008)
Finite Element Modeling of Resurfacing Hip Prosthesis: Estimation of Accuracy Through Experimental Validation
J Biomech Eng (February,2010)
Design Concept for a Total Knee Replacement with Condylar Guiding Features
J. Med. Devices (June,2011)
Related Proceedings Papers
Related Chapters
Novel and Efficient Mathematical and Computational Methods for the Analysis and Architecting of Ultralight Cellular Materials and their Macrostructural Responses
Advances in Computers and Information in Engineering Research, Volume 2
Elastic Constant Used in Continuum-Based Adhesion Models
International Symposium on Information Engineering and Electronic Commerce, 3rd (IEEC 2011)
Spring Constants of Elastic Shapes in Contact
Applying the ASME Codes: Plant Piping & Pressure Vessels (Mister Mech Mentor, Vol. 2)