In the past ten years there has been increased use of six-degree-of-freedom instrumented spatial linkages for the measurement of biological joint motions. In spite of the increased popularity, little information has been reported on the accuracy of these devices. In this paper, we present a two-part investigation of the accuracy of the instrumented spatial linkage when used to measure knee joint kinematics. In the first part, we present the results of a theoretical analysis and an experimental determination of the errors associated with spatial linkage systems. In the second part, we describe the errors associated with a bi-planar X-ray system used to obtain the coordinate transformation between the linkage ends and coordinate systems located in the bones comprising the joint. We found that the theoretical error analysis consistently overestimated the actual measurement error, thus providing an unreliable estimate of errors. The experimental study of both the linkage system and the bi-planar X-ray system demonstrated that the accuracy of displacement measurement is insensitive to large systematic errors in position measurement.
Skip Nav Destination
Article navigation
Research Papers
Error Analysis of a System for Measuring Three-Dimensional Joint Motion
W. J. Suntay,
W. J. Suntay
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
Search for other works by this author on:
E. S. Grood,
E. S. Grood
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
Search for other works by this author on:
M. S. Hefzy,
M. S. Hefzy
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
Search for other works by this author on:
D. L. Butler,
D. L. Butler
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
Search for other works by this author on:
F. R. Noyes
F. R. Noyes
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
Search for other works by this author on:
W. J. Suntay
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
E. S. Grood
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
M. S. Hefzy
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
D. L. Butler
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
F. R. Noyes
Giannestras Biomechanics Laboratory, Department of Orthopaedic Surgery, University of Cincinnati Medical Center, Cincinnati, Ohio 45267
J Biomech Eng. May 1983, 105(2): 127-135 (9 pages)
Published Online: May 1, 1983
Article history
Received:
December 7, 1981
Revised:
December 28, 1982
Online:
June 15, 2009
Citation
Suntay, W. J., Grood, E. S., Hefzy, M. S., Butler, D. L., and Noyes, F. R. (May 1, 1983). "Error Analysis of a System for Measuring Three-Dimensional Joint Motion." ASME. J Biomech Eng. May 1983; 105(2): 127–135. https://doi.org/10.1115/1.3138396
Download citation file:
Get Email Alerts
A Feasible Low-Cost System for Kinematic and Kinetic Analysis of Sit-to-Stand Movement
J Biomech Eng (April 2025)
Related Articles
Virtual Axis Finder: A New Method to Determine the Two Kinematic Axes of Rotation for the Tibio-Femoral Joint
J Biomech Eng (January,2010)
In Vitro Measurements of Porcine Anterior Column Units Under Free Swelling
J Biomech Eng (December,2003)
Related Proceedings Papers
Related Chapters
Some Effects of Experimental Error in Fracture Testing
Fracture Analysis: Proceedings of the 1973 National Symposium on Fracture Mechanics, Part II
A Six-Degree-of-Freedom Acoustic Transducer for Rotation and Translation Measurements Across the Knee
Skiing Trauma and Safety: Eighth International Symposium
Development of a Computer Vision Sensor for Structural Displacement Measurement
Computer Vision for Structural Dynamics and Health Monitoring